Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Graphtec AL3301/AL3302 (1 year) CAL VER /5520 DATE: 03-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 25 NUMBER OF LINES: 234 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # GRAPHTEC LINEARCORDER WR3310 MARK VII technical brochure. # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- R P F V 1.003 ASK+ X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP This procedure tests one AL3301 or AL3302 installed in 1.005 DISP any channel (AL3302 is double wide-80mm). 1.006 MEM2 Do you have an AL3301 or an AL3302? 1.007 JMPL AL3301 ZCMPI(MEM2, "AL3301") 1.008 JMPL AL3302 ZCMPI(MEM2, "AL3302") 1.009 DISP You must enter "AL3301" or "AL3302". 1.010 JMP 1.005 1.011 LABEL AL3301 1.012 MATH M[3] = 1 1.013 JMP 1.016 1.014 LABEL AL3302 1.015 MATH M[3] = 2 1.016 DISP Connect the 5520A to the [MEM2] as follows: 1.016 DISP [32] 5520A NORMAL HI to HI INPUT 1.016 DISP [32] 5520A NORMAL LO to LO INPUT 1.017 SET MAINFRAMEi : SETTINGS 1.017 SET POWER : ON 1.017 SET CHART EXT : OUT 1.017 SET HEAT : ON 1.017 SET CHART SPEED : 100 1.017 SET SEC / MIN : MIN 1.018 SET PLUG-IN : SETTINGS 1.018 SET MEAS/OFF/CAL.V : OFF 1.018 SET mV/x100mV : mV 1.018 SET RANGE : 1000 1.018 SET GAIN : FULL CW 1.018 SET POSITION : CENTER SCALE 1.019 DISP Set the proper channel HEAT switch to ON. 1.019 DISP If necessary, adjust the HEAT CONT. for best trace. 1.020 HEAD {BALANCE} 1.021 MATH MEM = .4 * M[3] 1.022 MESS While turning the RANGE control from maximum sensitivity 1.022 MESS to minimum sensitivity NOTE the pen shift from the 1.022 MESS center position. 1.022 MESS 1.023 EVAL Is the range balance pen movement <[MEM]mm? 2.001 RSLT =range balance pen movement <[MEM]mm 2.002 SET MEAS/OFF/CAL.V : MEAS 2.003 MESS Measure the pen movement while switching the 2.003 MESS MEAS/OFF/CAL.V between MEAS & OFF. 2.003 MESS 2.004 EVAL Is the measure balance pen movement <[MEM]mm? 3.001 RSLT =measure balance pen movement <[MEM]mm 3.002 HEAD {GAIN} 3.003 SET MEAS/OFF/CAL.V : MEAS 3.003 SET mV/x100mV : mV 3.004 MATH MEM = 500 / M[3] 3.005 DISP Set the RANGE : [MEM] 3.006 5520 1V S 2W 3.007 MATH MEM = 20 * M[3] 3.008 DISP Adjust the front panel GAIN control for a pen deflection 3.008 DISP of EXACTLY [MEM]mm. 3.009 5520 * S 3.010 HEAD GAIN: {1000mV RANGE} 3.011 DISP Set the RANGE : 1000 3.011 DISP 3.011 DISP Using the POSITION control, exactly center the trace. 3.012 MATH MEM = 2000 * M[3] 3.013 MATH MEM1 = 20 * M[3] 3.014 MESS SLEW the 5520 output for EXACTLY [MEM1]mm of deflection. 3.015 5520 mV 1% 2W 4.001 HEAD GAIN: {500mV RANGE} 4.002 DISP Set the RANGE : 500 4.002 DISP 4.002 DISP Using the POSITION control, exactly center the trace. 4.003 MATH MEM = 1000 * M[3] 4.004 5520 mV 1% 2W 5.001 HEAD GAIN: {200mV RANGE} 5.002 DISP Set the RANGE : 200 5.002 DISP 5.002 DISP Using the POSITION control, exactly center the trace. 5.003 MATH MEM = 400 * M[3] 5.004 5520 mV 1% 2W 6.001 HEAD GAIN: {100mV RANGE} 6.002 DISP Set the RANGE : 100 6.002 DISP 6.002 DISP Using the POSITION control, exactly center the trace. 6.003 MATH MEM = 200 * M[3] 6.004 5520 mV 1% 2W 7.001 HEAD GAIN: {50mV RANGE} 7.002 DISP Set the RANGE : 50 7.002 DISP 7.002 DISP Using the POSITION control, exactly center the trace. 7.003 MATH MEM = 100 * M[3] 7.004 5520 mV 1% 2W 8.001 HEAD GAIN: {20mV RANGE} 8.002 DISP Set the RANGE : 20 8.002 DISP 8.002 DISP Using the POSITION control, exactly center the trace. 8.003 MATH MEM = 40 * M[3] 8.004 5520 mV 1% 2W 9.001 HEAD GAIN: {10mV RANGE} 9.002 DISP Set the RANGE : 10 9.002 DISP 9.002 DISP Using the POSITION control, exactly center the trace. 9.003 MATH MEM = 20 * M[3] 9.004 5520 mV 1% 2W 10.001 HEAD GAIN: {5mV RANGE} 10.002 DISP Set the RANGE : 5 10.002 DISP 10.002 DISP Using the POSITION control, exactly center the trace. 10.003 MATH MEM = 10 * M[3] 10.004 5520 mV 1% 2W 11.001 HEAD GAIN: {5 x100mV RANGE} 11.002 DISP Set the RANGE : 5 11.002 DISP Set the mV/x100mV : [27][91]1mx100mV 11.002 DISP 11.002 DISP Using the POSITION control, exactly center the trace. 11.003 MATH MEM = 1 * M[3] 11.004 5520 V 1% 2W 12.001 HEAD GAIN: {10 x100mV RANGE} 12.002 DISP Set the RANGE : 10 12.002 DISP 12.002 DISP Using the POSITION control, exactly center the trace. 12.003 MATH MEM = 2 * M[3] 12.004 5520 V 1% 2W 13.001 HEAD GAIN: {20 x100mV RANGE} 13.002 DISP Set the RANGE : 20 13.002 DISP 13.002 DISP Using the POSITION control, exactly center the trace. 13.003 MATH MEM = 4 * M[3] 13.004 5520 V 1% 2W 14.001 HEAD GAIN: {50 x100mV RANGE} 14.002 DISP Set the RANGE : 50 14.002 DISP Set the mV/x100mV : [27][91]1mx100mV 14.002 DISP 14.002 DISP Using the POSITION control, exactly center the trace. 14.003 MATH MEM = 10 * M[3] 14.004 5520 V 1% 2W 15.001 HEAD GAIN: {100 x100mV RANGE} 15.002 DISP Set the RANGE : 100 15.002 DISP 15.002 DISP Using the POSITION control, exactly center the trace. 15.003 MATH MEM = 20 * M[3] 15.004 5520 V 1% 2W 16.001 HEAD GAIN: {200 x100mV RANGE} 16.002 DISP Set the RANGE : 200 16.002 DISP 16.002 DISP Using the POSITION control, exactly center the trace. 16.003 MATH MEM = 40 * M[3] 16.004 5520 V 1% 2W 17.001 HEAD GAIN: {500 x100mV RANGE} 17.002 DISP Set the RANGE : 500 17.002 DISP Set the mV/x100mV : [27][91]1mx100mV 17.002 DISP 17.002 DISP Using the POSITION control, exactly center the trace. 17.003 MATH MEM = 100 * M[3] 17.004 5520 V 1% 2W 18.001 HEAD GAIN: {1000 x100mV RANGE} 18.002 DISP Set the RANGE : 1000 18.002 DISP 18.002 DISP Using the POSITION control, exactly center the trace. 18.003 MATH MEM = 200 * M[3] 18.004 5520 V 1% 2W 19.001 MESS 19.002 HEAD {FREQUENCY RESPONSE} 19.003 SET RANGE : 100 19.003 SET mV / x100mV : mV 19.004 MATH MEM1 = M[3] -1.5 19.005 JMPT 22.002 19.006 MESS SLEW the 5520 for exactly 40mm p-p deflection. 19.006 MESS 19.007 5520 400mVpp 1H SI N 2W 19.008 5520 mVpp 60H SI S 2W 19.009 EVAL Is the {p-p deflection 40mm +/-.5mm(@60H)}? 20.001 MESS SLEW the 5520 for exactly 20mm p-p deflection. 20.001 MESS 20.002 5520 200mVpp 1H SI N 2W 20.003 5520 mVpp 100H SI S 2W 20.004 EVAL Is the {p-p deflection 20mm +/-.5mm(@100H)}? 21.001 MESS SLEW the 5520 for exactly 10mm p-p deflection. 21.001 MESS 21.002 5520 100mVpp 1H SI N 2W 21.003 5520 mVpp 140H SI S 2W 21.004 EVAL Is the {p-p deflection 10mm +/-.5mm(@140H)}? 22.001 JMP 25.001 22.002 MESS SLEW the 5520 for exactly 80mm p-p deflection. 22.002 MESS 22.003 5520 800mVpp 1H SI N 2W 22.004 5520 mVpp 30H SI S 2W 22.005 EVAL Is the {p-p deflection 80mm +/-1mm(@30H)}? 23.001 MESS SLEW the 5520 for exactly 40mm p-p deflection. 23.001 MESS 23.002 5520 400mVpp 1H SI N 2W 23.003 5520 mVpp 45H SI S 2W 23.004 EVAL Is the {p-p deflection 40mm +/-.5mm(@45H)}? 24.001 MESS SLEW the 5520 for exactly 20mm p-p deflection. 24.001 MESS 24.002 5520 200mVpp 1H SI N 2W 24.003 5520 mVpp 60H SI S 2W 24.004 EVAL Is the {p-p deflection 20mm +/-.5mm(@60H)}? 25.001 END